Products

PEA(Pseudomonas Exotoxin A) ELISA Kit

EU3133 Catalogue No.
Exotoxin A ELISA Kit, ETA ELISA Kit, NAD(+)--diphthamide ADP-ribosyltransferase ELISA Kit, PE ELISA Kit, eta ELISA Kit, PA1148 ELISA Kit, PEA ELISA Kit, Pseudomonas Exotoxin A ELISA Kit
Reactivity:
Pseudomonas aeruginosa
Range:
31.25-2000pg/ml
Sensitivity:
18.75pg/ml
Product Name
PEA(Pseudomonas Exotoxin A) ELISA Kit
Catalogue No.
EU3133
Alias
Exotoxin A ELISA Kit, ETA ELISA Kit, NAD(+)--diphthamide ADP-ribosyltransferase ELISA Kit, PE ELISA Kit, eta ELISA Kit, PA1148 ELISA Kit, PEA ELISA Kit, Pseudomonas Exotoxin A ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
PEA ELISA Kit allows for the in vitro quantitative determination of PEA concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
31.25-2000pg/ml
Sensitivity
< 18.75pg/ml
Species
Pseudomonas aeruginosa
UniProt ID
P11439
Standard curve
Storage
2-8 ℃ for 6 months
Recovery

Matrices listed below were spiked with certain level of PEA and the recovery rates were calculated by comparing the measured value to the expected amount of PEA in samples.

MatrixRecovery range(%)Average(%)
serum(n=5)87-10396
EDTA plasma(n=5)87-10094
heparin plasma(n=5)88-10495
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PEA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:8
serum(n=5)91-104%85-105%86-101%
EDTA plasma(n=5)83-97%84-94%84-93%
heparin plasma(n=5)80-98%81-100%81-93%
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note