Products

5-HIAA(5-Hydroxyindoleacetic Acid) ELISA Kit

EU2582 Catalogue No.
5-HIAA ELISA Kit, 5-Hydroxyindoleacetic Acid ELISA Kit
Reactivity:
Universal
Range:
1.563-100ng/ml
Sensitivity:
0.938ng/ml
Product Name
5-HIAA(5-Hydroxyindoleacetic Acid) ELISA Kit
Catalogue No.
EU2582
Alias
5-HIAA ELISA Kit, 5-Hydroxyindoleacetic Acid ELISA Kit
Detection Method
Competitive ELISA, Coated with Antigen
Application
5-HIAA ELISA Kit allows for the in vitro quantitative determination of 5-HIAA concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
1.563-100ng/ml
Sensitivity
< 0.938ng/ml
Species
Universal
Standard curve
Storage
2-8 ℃ for 6 months
Recovery

Matrices listed below were spiked with certain level of 5-HIAA and the recovery rates were calculated by comparing the measured value to the expected amount of 5-HIAA in samples.

MatrixRecovery range(%)Average(%)
serum(n=5)88-10095
EDTA plasma(n=5)85-9893
heparin plasma(n=5)86-10495
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 5-HIAA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:8
serum(n=5)89-103%87-102%90-95%
EDTA plasma(n=5)84-100%85-98%82-101%
heparin plasma(n=5)81-98%83-93%87-99%
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note
For Research Use Only