Products

Human BIRC7(Baculoviral IAP repeat-containing protein 7) ELISA Kit

EH2704 Catalogue No.
BIRC7 ELISA Kit, Livin ELISA Kit, KIAP ELISA Kit, ML-IAP ELISA Kit, baculoviral IAP repeat containing 7 ELISA Kit, baculoviral IAP repeat-containing 7 ELISA Kit, KIAPRING finger protein 50 ELISA Kit, Kidney inhibitor of apoptosis protein ELISA Kit, Livin ELISA Kit, livin inhibitor-of-apoptosis ELISA Kit, Melanoma inhibitor of apoptosis protein ELISA Kit, MLIAPlivin inhibitor of apoptosis ELISA Kit, ML-IAPmLiap ELISA Kit, RNF50LIVINbaculoviral IAP repeat-containing protein 7 ELISA Kit
Reactivity:
Human
Range:
125-8000pg/ml
Sensitivity:
75pg/ml
Product Name
Human BIRC7(Baculoviral IAP repeat-containing protein 7) ELISA Kit
Catalogue No.
EH2704
Alias
BIRC7 ELISA Kit, Livin ELISA Kit, KIAP ELISA Kit, ML-IAP ELISA Kit, baculoviral IAP repeat containing 7 ELISA Kit, baculoviral IAP repeat-containing 7 ELISA Kit, KIAPRING finger protein 50 ELISA Kit, Kidney inhibitor of apoptosis protein ELISA Kit, Livin ELISA Kit, livin inhibitor-of-apoptosis ELISA Kit, Melanoma inhibitor of apoptosis protein ELISA Kit, MLIAPlivin inhibitor of apoptosis ELISA Kit, ML-IAPmLiap ELISA Kit, RNF50LIVINbaculoviral IAP repeat-containing protein 7 ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
BIRC7 ELISA Kit allows for the in vitro quantitative determination of BIRC7 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
125-8000pg/ml
Sensitivity
< 75pg/ml
Species
Human
UniProt ID
Q96CA5
Standard curve
Storage
2-8 ℃ for 6 months
Recovery

Matrices listed below were spiked with certain level of BIRC7 and the recovery rates were calculated by comparing the measured value to the expected amount of BIRC7 in samples.

MatrixRecovery range(%)Average(%)
serum(n=5)85-9691
EDTA plasma(n=5)85-10095
heparin plasma(n=5)85-10597
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BIRC7 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:8
serum(n=5)89-95%91-99%87-105%
EDTA plasma(n=5)86-101%82-93%83-100%
heparin plasma(n=5)81-97%80-99%80-100%
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note