Products

Human CLCF1(Cardiotrophin-like cytokine factor 1) ELISA Kit

EH1866 Catalogue No.
CLCF1 ELISA Kit, BSF-3 ELISA Kit, BSF3 ELISA Kit, CISS2 ELISA Kit, CLC ELISA Kit, NNT-1 ELISA Kit, NNT1 ELISA Kit, NR6 ELISA Kit, B-cell stimulating factor 3 ELISA Kit, CRLF1 associated cytokine-like factor 1 ELISA Kit, neurotrophin-1 ELISA Kit, B-cell stimulating factor-3 ELISA Kit, novel neurotroph ELISA Kit, cardiotrophin-like cytokine factor 1 ELISA Kit, Novel neurotrophin-1 ELISA Kit
Reactivity:
Human
Range:
31.25-2000pg/ml
Sensitivity:
18.75pg/ml
Product Name
Human CLCF1(Cardiotrophin-like cytokine factor 1) ELISA Kit
Catalogue No.
EH1866
Alias
CLCF1 ELISA Kit, BSF-3 ELISA Kit, BSF3 ELISA Kit, CISS2 ELISA Kit, CLC ELISA Kit, NNT-1 ELISA Kit, NNT1 ELISA Kit, NR6 ELISA Kit, B-cell stimulating factor 3 ELISA Kit, CRLF1 associated cytokine-like factor 1 ELISA Kit, neurotrophin-1 ELISA Kit, B-cell stimulating factor-3 ELISA Kit, novel neurotroph ELISA Kit, cardiotrophin-like cytokine factor 1 ELISA Kit, Novel neurotrophin-1 ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
CLCF1 ELISA Kit allows for the in vitro quantitative determination of CLCF1 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
31.25-2000pg/ml
Sensitivity
< 18.75pg/ml
Species
Human
UniProt ID
Q9UBD9
Standard curve
Storage
2-8 ℃ for 6 months
Recovery

Matrices listed below were spiked with certain level of CLCF1 and the recovery rates were calculated by comparing the measured value to the expected amount of CLCF1 in samples.

MatrixRecovery range(%)Average(%)
serum(n=5)87-10497
EDTA plasma(n=5)88-10296
heparin plasma(n=5)90-10397
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CLCF1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:8
serum(n=5)87-96%89-105%90-104%
EDTA plasma(n=5)85-100%83-101%87-98%
heparin plasma(n=5)81-92%86-100%86-98%
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note
For Research Use Only