Products

Human MDC(Macrophage Derived Chemokine) ELISA Kit

EH0223 Catalogue No.
MDC ELISA Kit, CCL22 ELISA Kit, ABCD-1 ELISA Kit, MDC ELISA Kit, STCP-1 ELISA Kit, A-152E5.1 ELISA Kit, ABCD-1 ELISA Kit, CC chemokine STCP-1 ELISA Kit, C-C motif chemokine 22 ELISA Kit, chemokine ELISA Kit, C-C motif ligand 22 ELISA Kit, DC ELISA Kit, B-CK ELISA Kit, Macrophage-derived chemokine ELISA Kit, MDCStimulated T-cell chemotactic protein 1 ELISA Kit, MGC34554 ELISA Kit, SCYA22MDC ELISA Kit, 1-69 ELISA Kit, small inducible cytokine A22 ELISA Kit, small inducible cytokine subfamily A ELISA Kit, Cys-Cys ELISA Kit, member 22 ELISA Kit, Small-inducible cytokine A22 ELISA Kit, STCP-1 ELISA Kit, stimulated T cell chemotactic protein 1 ELISA Kit
Reactivity:
Human
Range:
62.5-4000pg/ml
Sensitivity:
37.5pg/ml
Product Name
Human MDC(Macrophage Derived Chemokine) ELISA Kit
Catalogue No.
EH0223
Alias
MDC ELISA Kit, CCL22 ELISA Kit, ABCD-1 ELISA Kit, MDC ELISA Kit, STCP-1 ELISA Kit, A-152E5.1 ELISA Kit, ABCD-1 ELISA Kit, CC chemokine STCP-1 ELISA Kit, C-C motif chemokine 22 ELISA Kit, chemokine ELISA Kit, C-C motif ligand 22 ELISA Kit, DC ELISA Kit, B-CK ELISA Kit, Macrophage-derived chemokine ELISA Kit, MDCStimulated T-cell chemotactic protein 1 ELISA Kit, MGC34554 ELISA Kit, SCYA22MDC ELISA Kit, 1-69 ELISA Kit, small inducible cytokine A22 ELISA Kit, small inducible cytokine subfamily A ELISA Kit, Cys-Cys ELISA Kit, member 22 ELISA Kit, Small-inducible cytokine A22 ELISA Kit, STCP-1 ELISA Kit, stimulated T cell chemotactic protein 1 ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
MDC ELISA Kit allows for the in vitro quantitative determination of MDC concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
62.5-4000pg/ml
Sensitivity
< 37.5pg/ml
Species
Human
UniProt ID
O00626
Standard curve
Storage
2-8 ℃ for 6 months
Recovery

Matrices listed below were spiked with certain level of MDC and the recovery rates were calculated by comparing the measured value to the expected amount of MDC in samples.

MatrixRecovery range(%)Average(%)
serum(n=5)88-10094
EDTA plasma(n=5)89-10596
heparin plasma(n=5)88-10296
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MDC and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:8
serum(n=5)91-102%86-97%85-102%
EDTA plasma(n=5)84-101%83-97%85-98%
heparin plasma(n=5)80-99%85-100%87-98%
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note