Products

Human Activin A ELISA Kit

EH0028 Catalogue No.
Activin A ELISA Kit, ACV-A ELISA Kit
Reactivity:
Human
Range:
15.625-1000pg/ml
Sensitivity:
9.4pg/ml
Product Name
Human Activin A ELISA Kit
Catalogue No.
EH0028
Alias
Activin A ELISA Kit, ACV-A ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
Activin A ELISA Kit allows for the in vitro quantitative determination of Activin A concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
15.625-1000pg/ml
Sensitivity
< 9.4pg/ml
Species
Human
Standard curve
Storage
2-8 ℃ for 6 months
Recovery

Matrices listed below were spiked with certain level of Activin A and the recovery rates were calculated by comparing the measured value to the expected amount of Activin A in samples.

MatrixRecovery range(%)Average(%)
serum(n=5)89-9992
EDTA plasma(n=5)92-10599
heparin plasma(n=5)86-10093
Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Activin A and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:8
serum(n=5)85-105%86-105%91-103%
EDTA plasma(n=5)89-94%86-101%85-93%
heparin plasma(n=5)89-100%85-93%87-96%
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note